MRI Research Today is a free monthly online journal that collates and summarizes the latest research about MRI, including details on magnetic resonance imaging, neuroimaging, brain tumors. | ||||||||
|
In vivo MRI using liquid nitrogen cooled phased array coil at 3.0 T.Kwok WE, You Z Department of Radiology, University of Rochester, Rochester, NY 14642, USA. edmund_kwok@urmc.rochester.edu A liquid nitrogen (LN2) cooled dual-channel array coil was designed and built for use on a 3.0-T whole-body scanner. In vivo imaging of a volunteer's fingers and imaging of a deceased mouse and oil phantom were performed using the LN2 cooled array and a similar room-temperature coil. Imaging results showed that the LN2 cooled array provides a signal-to-noise ratio gain of up to 240% as compared with its room-temperature counterpart. LN2 cooled arrays may be useful for high-resolution clinical imaging of joints, skin, eyes and peripheral vessels as well as for biomedical imaging of small animals in human disease modeling. Published 7 July 2006 in Magn Reson Imaging, 24(6): 819-23.
© 2005-2008 MRI Research Today. All Rights Reserved. |
| ||||||