Transceive surface coil array for MRI of the human prostate at 4T.Pinkerton RG, Near JP, Barberi EA, Menon RS, Bartha R Department of Medical Biophysics, University of Western Ontario, London, Ontario, Canada. A novel torso transceive surface coil array for prostate magnetic resonance imaging (MRI) and spectroscopy (MRS) at 4T is presented. It is shown that with the use of a conformal transceive surface coil array with 50 Omega transmitter amplifiers and receiver preamplifiers, one can perform whole-volume torso imaging while maintaining the high signal-to-noise ratio (SNR) inherent to surface coil designs. Recent theoretical considerations have shown that by focusing the infringing radiofrequency (RF) electromagnetic field, one can achieve increased penetration and signal homogeneity compared to a conventional circularly polarized driving scheme. A variation of this driving scheme particular to the proposed coil design resulted in a twofold increase in SNR in the prostate compared to that achieved with a conventional circularly polarized driving scheme. The novel transceive surface coil array presented is capable of full-volume imaging of the human torso at 4T while maintaining signal penetration in the deep region of the prostate gland. Published 5 February 2007 in Magn Reson Med, 57(2): 455-8.